Power Fail Alarm; High Pressure Alarm; High Leak Alarm - ResMed AutoSet CS 2 Clinician Manual

Hide thumbs Also See for AutoSet CS 2:
Table of Contents

Advertisement

W
ILL CLEAR WHEN
the low pressure condition is fixed
treatment is stopped by pressing the Start/Stop key.
P
OWER FAIL ALARM
T
RIGGERED BY
power failure
machine is disconnected or switched off while delivering treatment.
I
F A POWER FAIL ALARM SOUNDS
The flow generator stops delivering air pressure.
W
ILL STOP
when the Alarm Mute key is pressed, or
after 2 minutes, or
when power is restored.
CAUTION
!
Remove the mask from your face if the power fails.
H
IGH PRESSURE ALARM
T
RIGGERED BY
the mask pressure exceeds the set trigger level (25 cmH
milliseconds.
I
H
F A
1. The treatment will stop.
2. Turn power off.
3. Check that the air tube and sensor tubes are connected properly.
4. Turn power back on.
5. Remove mask and perform Air Circuit Learn function.
6. Try using the flow generator one more time.
7. If the high pressure alarm activates repeatedly, discontinue use and return to
ResMed for servicing. If the alarm does not recur, then continue to use as normal.
The alarm system is designed to ignore coughing. However, if a cough is especially intense
N
OTE
and prolonged, it may trigger the high pressure alarm.
W
ILL STOP WHEN
the A
H
IGH LEAK ALARM
T
RIGGERED BY
high mask leak (greater than 30 L/min) for more than 20 seconds.
I
F A HIGH LEAK ALARM SOUNDS
Adjust the mask to minimise leak.
54
:
:
:
:
IGH PRESSURE ALARM SOUNDS
:
S
CS2 is turned off.
UTO
ET
:
:
:
:
O) for more than 700
2