Cleaning And Maintenance; Daily; Weekly - ResMed AutoSet CS 2 Clinician Manual

Hide thumbs Also See for AutoSet CS 2:
Table of Contents

Advertisement

C
LEANING AND
You should regularly carry out the cleaning and maintenance described in this manual.
CAUTION
!
Do not wash the pressure sensor tube. If fluid enters the pressure sensor tube allow
it to dry completely by hanging it in a clean place out of direct sunlight. If the
pressure sensor tube cannot be dried completely it should be replaced.
D
AILY
1. Disconnect the air tubing and pressure sensor tube and hang them in a clean, dry
place until next use. Do not hang the air tubing in direct sunlight as it may harden
and crack over time.
2. Clean the mask according to the mask user instructions.
3. If you are using a humidifier, clean it according to the instructions in the manual.
W
EEKLY
1. Remove the air tubing and pressure sensor tube from the A
the mask.
2. Remove the pressure sensor tube and Proximal cuff from the air tubing.
3. Wash the mask system according to the instructions supplied with it.
4. Wash the air tubing and Proximal cuff in warm water using mild detergent. Rinse
thoroughly, hang and allow to dry.
5. Before next use, assemble the mask and headgear according to the mask user
instructions.
6. Reconnect the pressure sensor tube to the air tubing, then connect both to the
Proximal cuff. Attach the Proximal cuff to the mask.
Air tubing
Remove the pressure sensor tube
from air tubing
Figure 20: Undoing the air tubing assembly for cleaning
M
AINTENANCE
Remove Proximal cuff
Tubing clips
S
CS2 unit and
UTO
ET
C
M
LEANING AND
AINTENANCE
57