Instruction Manual - Clinic 3.40 Compressors

tRoubLESHootINg

faILuRE to foLLoW tHE pRoCEDuRES CoNtaINED IN tHIS paRagRapH
maY CauSE a maLfuNCtIoN of tHE maCHINE aND/oR uNDERmINE tHE
SafEtY of tHE SamE.
pRobLEm
The machine does
not turn or stops and
does not restart.
The machine does
not turn or stops and
does not restart.
The machine often
starts again without
the pneumatic plant
is operating.
18
MAN0340001 - 07/2019
CauSE
No electricity.
Intervention of the re-
settable thermal due to
current overload and/or
short circuit.
Air leaks from the pipe or
its pneumatic plant.
SoLutIoNS
• Check that the switch (18) is in pos.
"ON (I)".
• Check that the power cord plug (15)
is working properly and inserted into
the socket.
• Check that any extension cords used
and the current outlet are operatio-
nal.
• Check that the switch of the main po-
wer supply is functioning and in pos.
"ON (I)".
• Follow the procedure described in
par. "RESETTABLE THERMAL".
• Check that the pneumatic system is
not damaged.
• Check the integrity of the pipe con-
nections and air system.