Electric Connection - GENTILIN CLINIC 3.40 Series Instruction Manual

Table of Contents

Advertisement

Instruction Manual - Clinic 3.40 Compressors

ELECtRIC CoNNECtIoN

tHE maNufaCtuRER HaS No LIabILItY foR faILuRE oR maLfuNCtIoNS
of opERatIoN to tHE maCHINE CauSED bY VoLtagE DRopS EXCEEDINg
THE TOLERANCES PROVIDED BY THE DISTRIBUTING BODY (± 10% POWER
- FREQUENCY ± 2%).
tHE maINS to WHICH tHE maCHINE IS CoNNECtED muSt bE IN aCCoR-
DANCE WITH THE REQUIREMENTS OF THE LAWS OF THE COUNTRY OF USE,
mEEt tHE SpECIfICatIoNS CoNtaINED IN paR. "maCHINE Data" aND bE
pRoVIDED WItH a SuItabLE gRouNDINg SYStEm.
It IS CompuLSoRY to INStaLL upStREam of tHE poWER SuppLY LINE aN
ELECtRIC CuttINg DEVICE WItH a SuItabLE DISCoNNECtINg pRotECtIoN
CooRDINatED WItH tHE gRouNDINg pLaNt.
aNY tYpE of ELECtRIC matERIaL uSED foR tHE CoNNECtIoN muSt bE SuIta-
BLE FOR USE, MARKED "CE" IF SUBJECT TO LOW VOLTAGE DIRECTIVE 2006/95/
EC aND CompLYINg WItH tHE REQuIREmENtS of tHE REguLatIoNS IN foRCE
IN tHE CouNtRY of uSE of tHE maCHINE.
IF NECESSARY, IT IS COMPULSORY TO CONNECT THE MACHINE ONLY TO GENE-
RatoRS of ELECtRICItY WItH a poWER gREatER tHaN tHE poWER SuppLY
INSTALLED, TO SUPPORT THE ABSORPTION CURRENT UPON STARTUP.
faILuRE to foLLoW tHE WaRNINgS aboVE maY CauSE IRREVERSIbLE DamagE
to tHE ELECtRIC EQuIpmENt of tHE maCHINE aND tHE SubSEQuENt VoID of
tHE WaRRaNtY.
The machine can be connected to the mains by plugging
the power cord (15) into the power outlet (by the Custo-
mer).
MAN0340001 - 07/2019
15
27
www.gentilinair.com