Safety Devices - GENTILIN CLINIC 3.40 Series Instruction Manual

Table of Contents

Advertisement

Instruction Manual - Clinic 3.40 Compressors
SafEtY

SafEtY DEVICES

The Safety Devices adopted are as follows (for the references on the drawings, please see
the pages down):
REf.
DESCRIptIoN
1
Cooling fan protection
hood.
2
Resettable thermal.
3
Tank pressure gauge.
4
Safety valve.
5
Pressure switch.
It IS CompuLSoRY to CoNStaNtLY CHECk tHE gooD fuNCtIoNINg aND
tHE EffICIENCY of aLL SafEtY DEVICES IN tHIS maCHINE.
It IS CompuLSoRY to REpLaCE aNY SafEtY DEVICES maLfuNCtIoNINg aND/
oR DamagED.
28
www.gentilinair.com
It is a fixed guard that consists of a grid of plastic ma-
terial fixed with screws at the base of the twin cylinder
pumping unit. It prevents accidental contact with the
moving cooling fan.
It is a thermal protection installed on the electrical sy-
stem, which stops the electric motor in case of current
overload and/or short circuit.
Recovery can be done through the appropriate button.
It is a measuring device installed on the tank of the ma-
chine. It shows the air pressure inside the tank (0 ÷ 10
bar).
It is a certified safety valve (set at 10.5 bar), installed in
the pneumatic system under the pressure switch. It is
used to discharge the overpressure of the system if the
pressure switch, for any problems, does not function.
The intervention of the safety valve forces the operator
to turn off the machine and request the assistance of
the maintenance personnel (see par. "INTERVENTION
OF THE SAFETY VALVE").
It is an electro-pneumatic device (set at min. 6 bar, max.
8 bar) fitted onto the pneumatic plant.
It is used to control the automatic start of the machine
when the pressure drops to 6 bars and the automatic
stop when the pressure reaches 8 bars.
fuNCtIoN
MAN0340001 - 07/2019