Power Supply - it 3900e Owner's Manual

Light feedback regulated tungsten halogen source
Table of Contents

Advertisement

Innovative Photonics
An IEC 320 power inlet module is located on the rear panel.
internationally accepted power interface that can be used with a variety of power cords. The power
entry module contains the main power switch and the fuse box. The fuse holder allows the use of
dual 5x20mm fuses. In order to access the fuses, the power cord must be removed.
To change the fuses, start by removing the power cord. Insert a small flat blade screwdriver into
one of the notches near the socket and gently pry the door open. Pull the door straight out and
swing it into the socket (see Figure 2). Lift the fuse holder out of the housing. Replace the fuses
with T2.0 Amp, 5x20mm, size fuses. The fuses should be placed in the fuse holder as shown in
Figure 2. Lift the fuse holder out of the housing.
Replace fuses with T2.0 Amp, 5x20mm fuses. The fuses should be placed in the fuse holder as
shown in Figure 3. Replace the holder in the housing such that the fuses enter first. Swing the door
closed and snap it back into place.
Revision 2 09/01/16
Illumination
Technologies, Inc.
Fig 2 - IEC 320 Power Entry Module
Fig 3 - Fuse Arrangement