Aspect BIS VIEW Service Information Manual page 60

Monitoring system
Table of Contents

Advertisement

SECTION 7
________________________________________________________________________
Status Messages:
System error
[ 533, 538 ]
Export halted: USB
drive not detected
[ 539 ]
Export halted: USB
drive full
[ 540 ]
Export halted: BISx
not detected
[ 541 ]
Export halted: USB
drive full
[ 544 ]
Export halted: USB
drive not detected
[ 545 ]
DIAGNOSTICS AND TROUBLESHOOTING
Possible Causes:
(See Message 2000)
1. Removable drive is not
connected properly to
USB(A) port.
2. Drive is not formatted
properly.
3. Drive is incompatible or
defective.
4. USB port is defective.
1. Removable drive is not
connected properly to
USB(A) port.
2. Drive "write protect" is
locked.
3. Drive is incompatible or
defective.
4. Drive is full.
1. BISx disconnected.
2. Defective BISx cable.
3. Defective BISx.
4. Defective monitor.
(See Message 540)
(See Message 539)
7-10
Corrective Actions:
1. Check connection.
2. Format drive to FAT32.
3. Replace drive.
4. Replace monitor.
1. Check connection.
2. Verify "write protect"
switch on the drive is set to
"unlock" position.
3. Replace drive. Contact
your Aspect Technical
Service representative for a
list of compatible drives.
4. Replace drive.
1. Connect BISx. Verify all
cable connections.
2. Inspect/repair cable at
connector end.
3. Replace the BISx.
4. Replace monitor.

Advertisement