SECTION 4
_______________________________________________________________________
4. Using the Allen wrench supplied, secure pole clamp to monitor with the set screw
provided.
5. Replace black knob screw.
6. To attach to pole, place pole within clamp bracket and tighten screw using the black
finger knob.
4.2
Instrument Connections
Detailed connection instructions are provided in the BIS VIEW Operating Manual.
4.2.1 Connecting the BISx
1. Connect the BISx to the monitor
Holding the cylindrical connector with the flat side up, plug the BISx Monitor
Interface Cable into the BISx port on the front of the monitor.
Once connected, the BISx need not be disconnected again. However, if you wish to
disconnect the BISx cable from the monitor, carefully grasp the connector and pull.
DO NOT pull or twist the cable.
2. Connect the PIC to the BISx
Turn the 10-pin connector of the PIC so that the BIS logo is facing up, then plug it
into the BISx.
Once connected, the PIC need not be disconnected again, however, if you wish to
disconnect the PIC, grasp the connector housing and pull firmly. DO NOT twist or
pull apart by the cable wire.
4.2.2 Power Cord Connections
The BIS VIEW Monitoring System is designed to use only 3-conductor IEC hospital-grade
power cords. Cords must be type SJE, SJT, or SJO. Align power cord connector to power
cord port and insert firmly. Check for a firm connection.
4.3
Installation and Verification Procedure
1. Open packages and inspect for all components:
• Monitor
• Power cord
• Pole clamp
• BISx
• PIC (Patient interface cable, connects BISx to patient)
2. Connect power cable to monitor, plug power plug into appropriate wall outlet.
• Verify that light to right of ON/Standby button is yellow.
3. Start up monitor by pressing the ON/Standby button (lower right corner).
PREPARATION FOR USE AND INSTALLATION
Note:
4-4