 Permanently high blood pressure values can damage your
health and must be treated by your doctor!
 Always discuss your values with your doctor and tell him/her if
you have noticed anything unusual or feel unsure. Never rely
on single blood pressure readings.
 Under no circumstances should you alter the dosages of
drugs or initiate a treatment without consulting your doctor.
 Deviations between measurements taken by your doctor or in
the pharmacy and those taken at home are quite normal, as
these situations are completely different.
 The pulse display is not suitable for checking the
frequency of heart pacemakers!
 If you are pregnant, you should monitor your blood pressure
regularly as it can change drastically during this time.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.

Device care

Clean the device only with a soft, dry cloth.

Cleaning the cuff

Carefully remove spots on the cuff with a damp cloth and soap-
suds.
WARNING: Under no circumstances may you wash the
inner bladder!

Accuracy test

We recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). Please contact your
local Microlife-Service to arrange the test (see foreword).

Disposal

Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.

12. Guarantee

This device is covered by a 5 year guarantee from the date of
purchase. The guarantee is valid only on presentation of the guar-
antee card completed by the dealer (see back) confirming date of
purchase or the receipt.
 Batteries and parts that become worn with use are not included.
 Opening or altering the device invalidates the guarantee.
8
• The guarantee does not cover damage caused by improper
handling, discharged batteries, accidents or non-compliance
with the operating instructions.
• The cuff has a functional guarantee (bladder tightness) for 2 years.
Please contact your local Microlife-Service (see foreword).

13. Technical Specifications

Operating conditions: 10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity
Storage conditions: -20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
Weight:
415 g (including batteries)
Dimensions:
157.5 x 105 x 61.5 mm
Measuring procedure: oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic
Measurement range: 20 - 280 mmHg – blood pressure
30 - 199 beats per minute – pulse
Cuff pressure
display range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
4 x 1.5 V alkaline batteries; size AA
Mains adapter DC 6V, 600 mA (optional)
approx. 920 measurements
Battery lifetime:
(using new batteries)
Protected against solid foreign particles
IP20:
with a diameter of more than 12.5 mm,
no protection against water
ANSI/AAMI/IEC 80601-2-30;
Reference to
ANSI/AAMI/ISO 81060-2; IEC 60601-1;
standards:
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life: Device: 5 years or 10000 measurements
Cuff: 2 years or 10000 measurements
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
®