5. Turn the knob on the Product to adjust the display until it shows the same as the
DMM. After the output has become stable, push and hold the knob to continue. When
the readings are complete, the Product will go to the check mode.
6. When you are done with the check mode, push the knob to continue to the next
calibration point.

Voltage Measure

To calibrate the Voltage Measure mode:
1. Use the double banana jack cable to attach the current output of the Calibrator to the
voltage jacks of the Product.
2. After you have made the connections, if you are in voltage measure calibration mode,
the display should look as shown below.
3. Output 0 V from the Calibrator. After the output has become stable, push the knob to
continue. As the unit takes readings, the screen shows ------- .
4. Output 30 V from the Calibrator. After the output has become stable, push the knob
to continue. As the Product takes readings, the screen shows ------- . When the
readings are complete, the Product exits calibration mode and goes to volts mode.
The calibration adjust procedure is complete and the Product changes to normal
operation.
mA Loop Calibrator
Calibration Adjustment Procedure
hid006.eps
hid007.eps
hid008.eps
15