Page 2
5068 Ins-Cal Insulation Tester Calibration System This instrument provides all the equipment necessary to test and calibrate insulation test sets and megohm meters. It consists of: 1) 9 high precision selectable spot insulation resistances. 2) 4 low resistance spot values.
Page 4
5068 Ins-Cal Typical Operation Procedure 1) Fully charge the battery in the 5068. 2) Fully charge the battery in the EUT. (or fit fresh batteries). 3) Complete the set up procedure for the EUT. 4) Connect the 5068 leads to the EUT; (make certain the 4mm plugs on the free end are safe).
Page 5
5068 Ins-Cal Observations 1) Always remove the battery charger leads from both instruments before starting set 2) The high voltage test leads supplied are of a high quality and are rated at 5kV, use these leads where ever possible for all tests, except for low resistance or continuity tests, for these use the supplied 9542 4mm plug test lead set.
Please ensure that all units have details of the service required and all relevant paper- work including contact name, address and telephone number. Time Electronics Ltd Unit 11 Botany Industrial Estate Tonbridge, Kent, TN9 1RH Page 5...