Inficon ULTRATEST UL3000 Fab Translation Of The Original Operating Instructions page 93

Helium leak detector with sensor technology
Hide thumbs Also See for ULTRATEST UL3000 Fab:
Table of Contents

Advertisement

INFICON
Type
Notification
W300
Anode voltage too low
W301
Anode voltage too high
W302
Suppressor voltage too low
W303
Suppressor voltage too high
W304
Anode-cathode voltage too
low
W305
Anode-cathode voltage too
high
E306
Anode voltage wrong
W310
Cathode 1 broken
W311
Cathode 2 broken
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
Possible sources of
error
Short circuit of the
anode voltage
VI-Board or MSB
defective
MSB defective
Short circuit suppressor
or preamplifier
VI-Board or MSB
defective
MSB defective
Short circuit between
anode and cathode
VI-Board or MSB
defective
MSB defective
The anode voltage set
value is outside the
allowable range
MSB defective
Cathode defective
Connection to cathode
broken
VI-Board or MSB
defective
Cathode defective
Connection to cathode
broken
VI-Board or MSB
defective
Warning and error messages | 7
Remedy
• Contact customer service
• Contact customer service
• Contact customer service
• Contact customer service
• Contact customer service
• Contact customer service
• Contact customer service
• Contact customer service
• Contact customer service
• Perform a calibration
• Contact customer service
• Contact customer service
• Check that the other cathode has been
switched on
• Contact customer service
• Contact customer service
• Contact customer service
• Check that the other cathode has been
switched on
• Contact customer service
• Contact customer service
• Contact customer service
93 / 140

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ULTRATEST UL3000 Fab and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Ultratest ul3000 fab plus

Table of Contents