Installation

2.1 Assembly and Wiring

Depending on the customer selected output cable options, the DirectLine can be
wired to an appropriate 16-42 Vdc source using 2 different methods:
1) Cordset. See Figure 2-1.
2) Field wiring connector with customer supplied cable. See page 7.
Refer to Section 7 for wiring for CE Mark applications.
2.1.1 Cordset
Minimum: 16 V @ 250 ohms (max.)
Maximum: 42 V @ 1400 ohms (max.)
Chart Recorder,
Controller, etc.
(Use 250 ohms
for 1-5 V signal)
6
2. Installation
Power Supply
_ _
_
Load
Resistance
+
Black (–)
*Blue wire not used

Figure 2-1 Cordset Connection and Wiring

DirectLine® DL423 Sensor Module User Manual
+ +
Brown (+)
Shield
Wire
(Ground)
*
Cordset 6m (19.7')
Electronics
Module
9/11