APPENDIX B
12-LEAD ECG DATA TRANSMISSION TO
CATALYST™ MUSE® SYSTEM
The ZOLL M Series with 12-Lead and Catalyst™ MUSE® interface allows for direct transmission of 12-lead ECG
records to a GE Medical Systems Catalyst MUSE system. The Catalyst MUSE system provides access to online
patient ECG records and enables physicians to quickly view and compare prehospital or hospital 12-lead data.
Catalyst MUSE is a trademark of GE Medical Systems.
Prerequisites
Not all cell phones transmit to the Catalyst MUSE system reliably. See your local ZOLL Sales Representative for
a list of ZOLL-approved cell phones. Before you can transmit 12-lead data to a Catalyst MUSE system, you must
contact the Catalyst MUSE system administrator at the medical facility where the system is located. The system
administrator will then:
Authorize transmission to the system
Provide you with telephone numbers for accessing the system
Provide you with site and location identifiers for the M Series unit
Set up the M Series site and location identifiers on the Catalyst MUSE system
Configure the Catalyst MUSE system to forward ECGs to physicians or devices such as carts or fax
machines
Ask the person responsible for configuring the M Series unit to set up the telephone numbers and the MUSE site
and location identifiers that the Catalyst MUSE system administrator provided. It is recommended that the
M Series device identifier also be configured. For more information, see the following sections in the M Series
Configuration Guide, P/N 9650-0201-01:
Device Identifier
Fax/Communication Phone Numbers
MUSE Site and Location
When these activities are complete, you can begin transmitting 12-lead ECG data to the Catalyst MUSE system.
Note: Whenever possible, enter a patient identification number (ID) before transmitting 12-lead ECG data to a
Catalyst MUSE system. If the data does not have a unique ID number, the medical facility must search through all
of its unidentified records to find the patient's information. To set up a patient ID number, see the "Patient
Identification Number" section in this insert.