8. "FAX INTERRUPTED" or
"MUSE INTERRUPTED"
message is displayed.
9. "FAX NO CARRIER" or
"MUSE NO CARRIER"
message is displayed.
10. "FAX NO DIAL TONE" or
"MUSE NO DIAL TONE" message is
displayed.
11. "MODEM REQUIRED" message is
displayed.
12. "MUSE CANCEL" message is
displayed.
13. "MUSE DATA NOT VALID"
message is displayed.
9650-0215-01
1.
Power cycle M Series unit and retry transmission.
2.
Contact ZOLL Technical Service Department.
Receiving phone line is not picking up or the sending modem is
experiencing problems. The M Series automatically re-transmits
until the transmission is successful or the M Series operator
presses the Abort softkey.
If retransmissions fail:
1.
Check modem connection (card and extension cable) to
M Series.
2.
Power cycle cellular phone.
3.
If transmitting to a Catalyst MUSE system and multiple
Catalyst MUSE phone numbers are configured, press the
Abort softkey and transmit to an alternate number.
4.
If transmitting to a Catalyst MUSE system, check that the
Catalyst MUSE system's phone numbers are set up
correctly. See the ZOLL M Series Configuration Guide,
P/N 9650-0201-01, for further information.
1.
Check modem connection (card and extension cable) to
M Series.
2.
Check modem connection to phone jack or cellular phone.
3.
Power cycle cellular phone.
No PCMCIA modem card present. Insert PCMCIA modem card
in upper slot and repeat transmission.
The M Series operator pressed the Abort softkey or turned the
selector switch to Defib or Pacer.
An overriding activity, such as an alarm situation, preempted data
in the process of being saved to the M Series unit. The data is
incomplete or corrupted.
Re-acquire 12-lead data and re-transmit.
12-lead - 15