Daily Operational Verification
1.
Connect the V-lead cable to the 12-lead cable.
2.
Connect the lead wires of the 12-lead cable and V-leads to the patient or 12-lead simulator. If connecting to a patient place ECG
electrodes as indicated in the Electrode Placement Section.
3.
Connect the ZOLL 12-lead cable to the connector located on the rear of the M Series.
4.
Turn the Selector Switch of the M Series to Monitor.
5.
Select a normal sinus rhythm on the Simulator.
6.
Cycle through each of the 12 leads by pressing the LEAD button.
7.
Verify that good quality, artifact-free ECG signals are displayed and stabilized within 10 seconds on your M Series.
8.
Verify the ECG LEAD OFF message is NOT displayed on the screen.
9.
Verify the LOW BATTERY message is NOT displayed on the screen.
10. Verify the filter setting is set for 0.05 – 150 Hz (4x3).
11. Press and hold the RECORDER button for 3 seconds.
12. Verify appropriate ECG signals are printed. See 12-lead Summary Reports Section for reference.
13. Verify accuracy of time and date printed on the strip.
14. Visually inspect ECG Cable snaps for corrosion, particularly on the retaining wire inside the snap.
9650-0215-01
12-lead - 12