Fluke 7526A Calibration Manual page 41

Precision process calibrator
Hide thumbs Also See for 7526A:
Table of Contents

Advertisement

Calibrator
5522A
Step 17: 100 V Read Adjustment
Use the test equipment shown in Table 3. Connect the 5522A V source jacks to the
Product isolated jacks. Use the 8508A DMM to monitor the voltage. See Figure 17 for
the connections.
Enter 17 on the PC. Source three calibration points from 5522A:
0 V
50 V
100 V
For each calibration point, adjust the voltage for the maximum accuracy. Use the DMM
indication as feedback. Push ENTER on the PC when stable to go to the next calibration
point.
Calibration is complete. Enter 18 at the PC to exit calibration mode.
CALIBRATOR
Figure 17. 10 V Read Adjustment Connections
7526A
VOLTS/mA
LOOP
HART
SW TEST
PWR
Press for
Switch Test
VOLTS/mA
Reset
Press and Hold
for Switch Test
HI
INPUT
LO
100 V PK
MAX
Precision Process Calibrator
Calibration Adjustment
8508A
gxm022.eps
33