TC Type
Minimum
-50 °C
-25 °C
0 °C
100 °C
R
400 °C
600 °C
1000 °C
1600 °C
-50 °C
-25 °C
0 °C
100 °C
S
400 °C
600 °C
1000 °C
1600 °C
-250 °C
-200 °C
-100 °C
T
0 °C
200 °C
-200 °C
0 °C
U
200 °C
-200 °C
-100 °C
XK
0 °C
600 °C
0 °C
200 °C
600 °C
BP
800 °C
1600 °C
2000 °C
[1]
Does not include thermocouple wire error.
Type B, E, J ,K, N, R, S and T are based on ITS-90
Type L and U are based on DIN 43710-1985
Type C is based on ASTM standard E 988-96
Type XK and BP are based on GOST R 8.585-2001
Range (°C)
Maximum
-25 °C
0 °C
100 °C
400 °C
600 °C
1000 °C
1600 °C
1767 °C
-25 °C
0 °C
100 °C
400 °C
600 °C
1000 °C
1600 °C
1767 °C
-200 °C
-100 °C
0 °C
200 °C
400 °C
0 °C
200 °C
600 °C
-100 °C
0 °C
600 °C
800 °C
200 °C
600 °C
800 °C
1600 °C
2000 °C
2500 °C
Precision Process Calibrator
Absolute Uncertainty, tcal ±5 °C, ±(°C)
Output/Input
90 days
1 Year
0.54 °C
0.55 °C
0.44 °C
0.45 °C
0.38 °C
0.39 °C
0.27 °C
0.28 °C
0.21 °C
0.22 °C
0.19 °C
0.21 °C
0.18 °C
0.19 °C
0.21 °C
0.23 °C
0.51 °C
0.51 °C
0.43 °C
0.43 °C
0.37 °C
0.38 °C
0.28 °C
0.29 °C
0.22 °C
0.23 °C
0.21 °C
0.22 °C
0.20 °C
0.22 °C
0.24 °C
0.26 °C
0.34 °C
0.35 °C
0.14 °C
0.16 °C
0.09 °C
0.11 °C
0.07 °C
0.09 °C
0.06 °C
0.09 °C
0.15 °C
0.16 °C
0.08 °C
0.10 °C
0.07 °C
0.10 °C
0.10 °C
0.11 °C
0.07 °C
0.09 °C
0.06 °C
0.08 °C
0.07 °C
0.09 °C
0.17 °C
0.18°C
0.14 °C
0.16 °C
0.15 °C
0.17 °C
0.22 °C
0.23 °C
0.26 °C
0.28 °C
0.38 °C
0.40 °C
Specifications