Fluke 7526A Calibration Manual page 29

Precision process calibrator
Hide thumbs Also See for 7526A:
Table of Contents

Advertisement

To test the ohm input:
1. Apply a four-wire short to the Product 4W RTD/Ω input.
2. For the first test, set the Product to the 400 Ω range.
3. For the second test, connect the Product 4W RTD/Ω input to the output of the 5522A
as shown in Figure 8.
4. Set the Product to the 400 Ω range. Note the indication in Table 8.
5. Set the Product to the 4000 Ω range after the 400 Ω test. Note the indication in
Table 8.
6. Output the same nominal values and note the Product indication in the 7526A
Reading column of Table 8.
7. Subtract the 5522A Actual Value from the 7526A Reading and ensure that it is
within the specified tolerance.
Calibrator
5522A
CALIBRATOR
Figure 8. Ohms Measurement Performance Test Connections
Precision Process Calibrator
Performance Tests
7526A
7526A
PRECISION PROCESS CALIBRATO
mA
VOLTS
RTD/
100 mA MAX
OUTPUT
HI
HI
OUTPUT
100V MAX
LO
LO
20 V PK
MAX
4 W RTD/
INPUT
HI
CURRENT
SENSE
LO
20 V PK
MAX
gxm016.eps
21