Agilent Technologies InfinityLab LC Series User Manual page 109

Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

8
Test Functions
Refractive Index Calibration
Preparing the Calibration
When
Recommended after exchange of the optical unit or RIM board.
Tools required
Description
Laboratory balance
Parts required
#
1
1
1
1
1
1 Preparation of the sucrose calibration solution.
2 Preparing the pump.
3 Flush the degasser and pump.
Agilent InfinityLab LC Series RID User Manual
p/n
Description
DAB/Ph Eur/BP/JP/NF/USP Grade Sucrose
9301-1446
Syringe
9301-0407
Needle
5190-5111
Syringe filter, 0.45 µm, 100/pk
0100-1516
Fingertight fitting PEEK, 2/pk
a To prepare 25 ml of the calibration solution 87.5 mg of the Sucrose
sample is required.
b Add the weighed amount of sample into a suitable volumetric flask.
c Dispense 10 ml of LC grade water into the flask and shake or stir to
dissolve.
d Dilute the contents of the flask to volume with LC grade water.
Wait five minutes and shake again. The solution is now ready for use.
a Fill a suitable solvent bottle with LC grade water.
b Connect this bottle to Channel A of the pump, A1 if a binary pump.
109