Dräger EPAS Operating Manual page 17

Table of Contents

Advertisement

Alcotest 7410
Plus
Error Code
Cause
E6
Instrument Parameter
Error
E7
Instrument Parameter
Error
E8
Timing malfunction
Remedy
Contact Draeger Safety,
Inc., Breathalyzer
Division, Technical
Department.
Contact Draeger Safety,
Inc., Breathalyzer
Division, Customer
Support Department.
Instrument will power
down after 10 seconds.
Turn back on.
If problem persists,
contact Draeger Safety,
Inc., Breathalyzer
Division, Customer
Support Department.
30
Printer
Fault
Cause
Not printing although
Voltage too low, too
supply voltage is
high, or wrong polarity
supplied.
Defective Fuse
Malfunction
Although supply
Faulty power cord
voltage is available,
paper feed is irregular or Power supply
printer stops.
inadequate or faulty
Mounting socket broken
Message Printed
Cause
TRANSMISSION
Incorrect log received
ERROR
HARDWARE ERROR
Hardware error
Remedy
Detach unit from power
supply, provide voltage
greater than 9.5 V/1A
and less than 17V/1A.
Detach unit from power
supply and replace fuse
(1A).
Contact Draeger Safety,
Inc., Breathalyzer
Division, Customer
Support Department.
Replace power cord.
Provide voltage greater
than 9.5V/1A and less
than 17V/1A.
Contact Draeger Safety,
Inc., Breathalyzer
Division, Customer
Support Department.
Remedy
Position Alcotest
7410
®
again, or clean
Plus
transparent window or
lenses.
Contact Draeger Safety,
Inc., Breathalyzer
Division, Customer
Support Department.
31

Advertisement

Table of Contents
loading

Table of Contents