Editing Operator or Subject Data
You can edit the Operator or Subject data by tapping the data field
you want to edit, or you can follow the steps below to edit the Opera-
tor or Subject Data.
1.) Select "Edit Op." or "Edit Sub." from the Cassiopeia
2.) The data entry sequence will be repeated, allowing you to make
the neccessary changes.
3.) When all data has been verified, the Cassiopeia
"Verify Data " screen and allow you to initiate the breath test.
4.) When the data is correct, select "Data OK" from the touchscreen
to send the data to the Alcotest
After selecting "Data OK", the data will be sent to the Alcotest
If the data is successfully transferred, a series of three beeps will
sound from the Alcotest
7410
®
has been initiated. The Alcotest
the "Choose Option" screen will appear on the Cassiopeia
tion for the next subject.
Evidential Test Warm Up Sequence
The Alcotest
7410
will start the evidential test warm up sequence
®
Plus
by displaying the date:
month
followed by the time:
hour
followed by:
followed by:
(Air Blank result)
touch screen.
®
will return to the
®
7410
.
®
Plus
7410
®
indicating that the testing mode
Plus
7410
will begin to warm up and
®
Plus
in prepara-
®
day
year
minute
(Air Blank)
16
Note: If alcohol is detected by the Alcotest
After a successful Air Blank, the Alcotest
then
the horn sounding and the green "READY" light illuminating.
Performing a Breath Test
1.) Remove the Alcotest
.
Plus
the mouthpiece holder.
2.) Attach a mouthpiece, ensuring that the
arrows on the mouthpiece and mouthpiece
holder point the same direction. Align the
ridges on the mouthpiece with the grooves
on the mouthpiece holder making sure
the mouthpiece fit is snug against the
mouthpiece holder.
3.) Instruct the subject to take a deep breath and blow evenly as long
as the horn sounds.
4.) When the subject has successfully provided a breath sample,
the result will automatically be displayed on the LCD display.
(Display time of the results will vary depending on the alcohol
concentration.)
After displaying the breath test result for five seconds, the Alcotest
7410
Plus
breath test can begin.
Note: If the subject does not fulfill the volume and/or flow rate
When
ent air, or the Alcotest
7410
®
from the previous breath test, the Alcotest
automatically perform additional Air Blanks until the
results are 0.000.
(Breath Test Number 1), followed simultaneously by
7410
from its cradle by pulling up on
®
Plus
will initiate a 120 seconds count down before the second
requirements,
will be displayed, indicating an
insufficient breath sample. To complete the test, the subject
must repeat the breath sample after the green "READY"
light comes on.
is displayed, repeat the steps above.
17
7410
in the ambi-
®
Plus
has not reached 0.000
Plus
7410
will
®
Plus
7410
will display
®
Plus
�������� ����
����
®