Maintenance and Troubleshooting
The maintenance and troubleshooting tables that follow are provided to
help eliminate confusion and prevent downtime by supplying corrective
procedures.
If problems persist, call Draeger Safety, Inc., Breathalyzer Division,
Support and Technical Service at (970) 385-5555.
Alcotest 7410
Plus
FAULT
CAUSE
Instrument switches off
Battery pack is
automatically or yellow
discharged, or battery
light is not lit after
pack is completely
switching on, or is only
exhausted.
faintly lit; the display
shows incomplete
symbols.
"READY" light is not lit Instrument malfunction
after switching on and 4
minute waiting interval
has not elapsed.
No continuous audible
Instrument not yet ready
tone when blowing into
for measurement. Horn
the instrument.
needs replacement (very
unusual) or intake is
blocked by foreign matter. Safety, Inc.,
Subject not blowing hard Breathalyzer Division
enough or mouthpiece
not seated properly.
REMEDY
Recharge battery pack
or replace with another
power pack.
Contact Draeger Safety,
Inc., Breathalyzer
Division, Customer
Support Department.
Have subject blow
harder or check the
mouthpiece.
Contact Draeger
Customer Support
Department.
28
Alcotest 7410
Plus
Error Code
Cause
Acc Err
Accuracy Check is out
of limits.
INS
Test subject is not
blowing hard enough or
steadily enough. Intake
is blocked by foreign
matter. Pressure sensor
or flow rate suspect.
rFi
RFI present.
E2
Instrument does not
recognize calibration
clip. Cal switch
malfunction.
E3
Calibration data invalid,
lifetime of sensor
reached or cal data
malfunction.
E4
Improper value was used
for calibration.
E5
Fault in sampling system
(due, most likely, to
motor malfunction).
Remedy
Redo Accuracy Check
and Calibrate if necessary.
Ask subject to blow
harder or more steady.
Sensor or flow rate
must be checked by
Draeger Safety, Inc.,
Breathalyzer Division,
Technical Department.
Change location or
remove RFI source.
Attach calibration clip
correctly before turning
on instrument. Cal
switch must be replaced
by an authorized
Service Technician.
Calibrate instrument.
Sensor must be replaced
or inspected by Draeger
Safety, Inc.,
Breathalyzer Division.
Calibrate with proper
value.
Repeat measurement. If
E5 reappears, contact
Draeger Safety, Inc.,
Breathalyzer Division.
29