1 Turn on the meter and select °C mode by pressing MODE if necessary.
Disconnect the temperature probe.
2 Press CAL to begin temperature calibration mode. "CA" (calibrate) will
display briefly. The factory default temperature 25.0°C value will blink.
3 Use MODE/INC to adjust the reading to the desired value. The maximum
adjustment is ± 0 to 100°C.
4 Press HOLD/ENTER to confirm the value. "CO" (confirm) will display
briefly. The meter will automatically save the calibration, cease blinking,
and begin °C measurement with the new default temperature.
Calibration
3
13