Introduction; Scope Of This Manual; Safety - Edwards RV Series Instruction Manual

Pump cartridge kits
Table of Contents

Advertisement

1

Introduction

1.1

Scope of this manual

This manual provides installation instructions for the cartridge kits for the Edwards RV rotary pumps. You must use
the cartridge kits as specified in this manual.
Use a cartridge kit to replace the cartridge, springs, seals and fluoroelastomer components in an RV rotary pump. We
recommend that you use all of the components in the cartridge kit when you service the pump.
The contents of the kit are defined in
1.2

Safety

Do not touch or inhale the thermal breakdown products of fluorinated materials which may be
present if the pump has been heated to 210 °C and above. These breakdown products are very
dangerous. Some of the seals in the pump are made from fluorinated materials.
The dynamic seals and O-rings used in this pump are made from fluorinated materials. Fluorinated materials are safe
in normal use but can decompose into very dangerous materials (which may include hydrofluoric acid) if they are
heated to 210 °C and above.
The pump may have overheated if it was misused or if it was in a fire. If the pump has overheated, take extreme care
to avoid skin contact with any part of the pump and to avoid inhalation of vapours from the pump. Health and Safety
Data sheets for fluorinated materials used in the pump are available on request: contact your supplier or Edwards.
© Edwards Limited 2011. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Section
2.
Section 3
describes how to install the kit components in the pump.
WARNING
A653-01-840 Issue F
Page 1

Hide quick links: