Option EXM, External Mixing Upgrade Kit
Analyzer Disassembly
If the instrument is placed on its face during any of the following procedures,
CAUTION
be sure to use a soft surface or soft cloth to avoid damage to the front panel,
keys, or input connector.
Make sure any adapters on the front panel are removed.
NOTE
If the analyzer has Option PRC, Portable Configuration, refer to the "Option
NOTE
PRC." section on page 10 to remove the outer case.
Standard Instrument
1. Disconnect the instrument from ac power.
2. Refer to Appendix Figure 1, "Instrument Outer Case Removal,". Using the T-20 driver, remove
the 4 screws (1) (two on each side) that attach the handle strap (2) on each side of the
instrument.
3. Remove the four key locks from the instrument four bottom feet.
4. Remove the four instrument bottom feet.
5. Using the T-20 driver, remove the four screws (including washers) (3) that hold the rear feet
(4) in place.
6. Pull the instrument cover (5) off towards the rear of the instrument.
8
Installation Note N9010-90041