Download Print this page

Keysight Technologies EXA Installation Note page 7

Signal analyzer option exm, external mixing upgrade kit
Hide thumbs Also See for EXA:

Advertisement

2. After the analyzer has completed turning on, press System, Show, System. Make note of the
following information from the Show System screen:
Product Number: _____________
Serial Number: ______________
Instrument S/W Revision: ______________
3. Check for the presence of one of the following frequency range options listed below in the
Show System. Put a check mark or "X" after the frequency range option listed below that
appears in the Show System menu.
N9010A-503_____
N9010A-507_____
N9010A-513_____
N9010A-526_____
N9010A-532_____
N9010A-544_____
4. If the analyzer is equipped with any of the following frequency range options (refer to the data
in step 3 above):
N9010A-503
N9010A-507
N9010A-513
N9010A-526
do not proceed with the installation of this upgrade kit.
An RF or microwave EXA is not compatible with Option EXM.
Installation Note N9010-90041
Option EXM, External Mixing Upgrade Kit
7

Advertisement

loading