Prn 50M; Cleaning And Disinfection Overview; Cleaning; Cleaning Warnings - GE CARESCAPE PDM Supplement Manual

Cleaning and disinfecting
Table of Contents

Advertisement

PRN 50M+

Cleaning and disinfection overview

The following cleaning and disinfection information applies to the PRN 50M+.
The information provided in this chapter does not supersede any instructions for use
provided by the manufacturer or provided with the device.
For cleaning, disinfection, and care information for accessories made by other
manufacturers, see the applicable instructions for use provided by the manufacturer.

Cleaning

Cleaning warnings

WARNING
WARNING
WARNING
WARNING
WARNING
WARNING

Permitted detergents

Water
Enzymatic or neutral pH detergents
2110145-001
Avoid using chemicals other than the ones described in this
document, as they may damage device surfaces or labels,
or cause equipment failures.
If liquid has accidentally entered the system or its parts,
disconnect the power cord from the power supply, remove
the battery, and have the equipment serviced by qualified
service personnel.
Never immerse any part of the device, cables, or leadwires in
liquids or allow liquid to enter the interior of the device.
Do not autoclave any part of the system with steam or sterilize
with ethylene oxide.
Do not pour or spray any liquid so that it may seep into
connections or openings.
Never use conductive solutions, oxidizing compounds, wax, or
wax compounds to clean device.