Cleaning Non-Applied Parts, General Instructions; Other Devices' Cleaning Instructions; Disinfection; Disinfection Points To Note - GE CARESCAPE PDM Supplement Manual

Cleaning and disinfecting
Table of Contents

Advertisement

Cleaning non-applied parts, general instructions

Follow these instructions to clean the monitor, modules, EEG headbox, and other
non-applied parts unless there are separate part-specific instructions. Non-applied
parts refer to those parts of the system that are not in direct contact with the patient.
Be especially careful when cleaning the display or monitor screen. They are more
sensitive to rough cleaning methods than for example the monitor housing.
1. Turn off the power to the equipment.
2. Disconnect the equipment from the power supply.
3. Remove all cables and batteries (if applicable) and close battery door(s).
4. Dampen a soft lint-free cloth with one of the permitted detergents.
5. Wring excess liquid from the cloth and wipe the exterior surface. Pay special
attention to hard to clean areas, like grooves and crevices.
Any contact of cleaning or disinfecting solutions with metal parts may cause
corrosion.
Do not damage or bend connector pins when cleaning or drying.
6. Remove any soil as soon as possible after use by wiping the device.
Do not let fluid pool around connection pins. If this happens, blot dry with a cotton
swab or soft cloth.
7. Wipe off the cleaning solutions with a clean, lightly moistened cloth.
8. Dry thoroughly with a dry, lint-free cloth and let air dry for at least 30 minutes.
Drying times may vary based on the environmental conditions.
9. Visually ensure that the device is clean. If visible soil remains, repeat cleaning
procedure until the device is thoroughly clean.
10. Insert batteries (if applicable) and close battery doors.
11. Reconnect the equipment to the power supply.
12. Turn on the power to the equipment.

Other devices' cleaning instructions

For details about cleaning the accessories, see the instructions for use in the
accessory package.
Do not reuse single-use disposable accessories.
Refer to the user documentation provided with the keyboard, mouse, and barcode
reader for instructions on how to clean them. Always consider your hospital guidelines
as well.

Disinfection

Disinfection points to note

Always clean before disinfecting.
2110145-001
Cleaning and disinfecting
CARESCAPE B650 Monitor
51