Other Manufacturer Cleaning Instructions; Disinfection; Disinfection Points To Note; Permitted Disinfectants - GE CARESCAPE PDM Supplement Manual

Cleaning and disinfecting
Table of Contents

Advertisement

5. Wring excess liquid from the cloth and wipe the exterior surface. Pay special
attention to hard-to-clean areas, like grooves and crevices.
Remove any soil as soon as possible after use by wiping the device.
Any contact of cleaning solutions with metal parts may cause corrosion.
Do not damage or bend connector pins when cleaning or drying.
Do not let fluid pool around connection pins. If this happens, blot dry with
a cotton swab or soft cloth.
6. Wipe off the cleaning solutions with a clean, lightly moistened cloth.
7. Visually ensure that the device is clean. If visible soil remains, repeat cleaning
procedure until the device is thoroughly clean.
8. Dry thoroughly with a dry, lint-free cloth and let air dry for at least 30 minutes.
Drying times may vary based on the environmental conditions.
9. Reconnect the device to the power supply.
10. Reconnect all cables.
11. Turn on the power to the device.

Other manufacturer cleaning instructions

For details about cleaning and disinfection information for devices, device
components, supplies, and accessories made by manufacturers other than GE, see
the applicable instructions for use provided by the manufacturer.

Disinfection

Disinfection points to note

Always clean before disinfecting.
Always dilute cleaning and disinfectant agents according to their manufacturer's
instructions. Always consider your hospital guidelines as well.
Use only the permitted substances.
Your hospital guidelines permitting, all cleaning and disinfection activities can
be carried out at the bedside.
Before drying a surface after wiping it, wait the minimum time required according
to the substance manufacturer's instructions.
Visually ensure that no substance residue remains on the device.

Permitted disinfectants

About the following table:
All third party trademarks are property of their respective owners.
Trademark names and product availability may vary by country. Consult the
column that lists ingredients to determine if an equivalent disinfectant is available
in your country.
2110145-001
Cleaning and disinfecting
CARESCAPE Central Station
33

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents