SPECIFICATIONS

Specifications subject to change without notice.
Oxygen Measurement:
Range
Accuracy
Temperature
Resolution:
Oxygen Concentration
% Saturation
Temperature
Display: Custom LCD
Inputs:
5-pin shielded
circular connector
Power Requirements: 6–12 Vdc; use either Hach-supplied 115 or 230V, 50/60 Hz external
power supply or a customer-provided supply with 50 mA output, 5.5-mm power plug with a
2.5 mm center post opening
Installation Category: II
Instrument drift: < 1%/day
Environmental requirements: 5 to 50 °C at 85% relative humidity, non-condensing
Meter dimensions: 15 x 25.4 x 8.37 cm (10.15 x 6 x 3.5 inches)
Enclosure: Water resistant (IP32), chemical resistant
Probe Dimensions (Model 50180):
Probe Length
Body Diameter
Cap
Cable Length
Connector
0-20 mg/L (ppm)
0-200% sat.
± 1% full scale
0-50 °C
0.01 or 0.1 ppm (mg/L)
0.1%
0.1 °C
Outputs:
RS232C
150 mm
12 mm
15 mm dia. x 35 mm
1, 3, or 15 meters
Fisher