Installation Procedure Over Usb - Keysight Technologies PXA N9030A Installation Note

Signal analyzers, option lnp low noise path or option mpb microwave preselector bypass
Hide thumbs Also See for PXA N9030A:

Advertisement

Contents
Quantity
Description
1
Installation Note
1
Option Upgrade Entitlement Certificate
Tools Required
— Personal computer with internet access and USB port
— USB storage device with > 2 GB free memory
If adjustments and performance testing is required:
— Keysight Calibration and Adjustment Software, N7814A
— Test equipment and computer supported by the X-Series Performance Tests and Adjustment
Software.
Optional tools:
— Instrument Service Guide - purchase as N9030AK-OBW or N9030-90015

Installation Procedure over USB

1. Locate the Option Upgrade Entitlement Certificate (5964-5178) from the kit.
2. Redeem the Option Upgrade Entitlement Certificate by following the instructions on the
Certificate.
3. After redeeming your Option Upgrade Entitlement Certificate you will receive an email with an
attached License File.
4. Locate a USB storage device. Perform a virus scan on this device before use.
5. Save the License File to the root directory of the USB storage device.
6. Connect the USB storage device to the signal analyzer USB port. Windows will detect the new
hardware and may display the configuration menu shown in
configured according to your preferences.
Installation Note N9020-90181
Option LNP or Option MPB Installation Kit
Keysight Part
Number
This note
5964-5178
Figure
1. This menu may be
5