To test the single-clock, single-edge, state acquisition

Testing the single-clock, single-edge, state acquisition verifies the performance of
the following specifications:
• Minimum master-to-master clock time.
• Maximum state acquisition speed.
• Setup/Hold time for single-clock, single-edge, state acquisition.
This test checks two combinations of data channels using a single-edge clock at
two selected setup/hold times.
Equipment Required
Equipment
Pulse Generator
Digitizing Oscilloscope
Adapter
SMA Coax Cable (Qty 3)
Coupler
BNC Test Connector,
6x2 (Qty 4)

Set up the equipment

If you have not already done so, do the following procedures:
"To set up the test equipment and the logic analyzer" on page 23.
"To set up the logic analyzer for the state mode tests" on page 33.

Connect and configure the logic analyzer

1 Using the 6-by-2 test connectors, connect the first combination of logic
analyzer clock and data channels listed in one of the following tables to the
pulse generator.
If you are testing a 1680/81/90/91A,AD, you will repeat this test for the second
combination.
To test the single-clock, single-edge, state acquisition
Critical Specifications
200 MHz 2.5 ns pulse width, <600 ps rise time 8133A option 003
≥6 GHz bandwidth, <58 ps rise time
SMA(m)-BNC(f)
18 GHz bandwidth
BNC(m)(m)
Chapter 3: Testing Performance
Recommended Model/Part
54750A w/ 54751A
1250-1200
8120-4948
1250-0216