On-Board FastRegen Control Module Programming and Operation Instructions
Table B-1: On-Board Module RS-232 Commands (Continued)
Command
P
[5 digit number]
P/N 8040406
Parameter
[0 - 6,A,G,S]
Reads/Sets a regeneration parameter.
| [?]
0 = Pump restart delay (minutes) range = 0 - 59994
1 = Extended purge time (minutes) range = 0 - 9999
2 = Repurge cycles 0 - 20
3 = Rough to pressure range = 25 - 200
4 = Rate-of-rise (microns/minute) range = 1 - 100
5 = ROR cycles range = 0 - 40
6 = Start-up temperature (after power failure) range = 0 - 80
A = Clear/set rough valve interlock range = 0 - 1
G = Repurge time (minutes) range = 0 - 9999
S = Fast rough test range = 10 - 999
C
TI
HELIX TECHNOLOGY CORPORATION
Description
-
CRYOGENICS
B-7