Roughing Valve; Delay Activation Of Relay - HELIX CTI-CRYOGENICS On-Board FastRegen Programming And Operation Instructions

Control module
Table of Contents

Advertisement

On-Board FastRegen Control Module Programming and Operation Instructions

Roughing Valve

Delay Activation of Relay

P/N 8040406
The tenth parameter of the RELAYS function allows you to program the
setpoint relay to activate when the rough valve is opened and deactivate
when the rough valve is closed as shown in Figure 3-57.
Last
3
1
2
Next
4
6
5
Clear
7
8
9
-
0
.
Figure 3-57: Roughing Valve Display
The eleventh parameter of the RELAYS function allows you to establish a
delay time during which your selected relay waits before operating as
shown in Figure 3-58. This parameter will delay both activation and
deactivation of the relays.
Last
3
1
2
Next
4
5
6
Clear
9
7
8
-
0
.
Figure 3-58: Delay Activation of Relay Display
ROUGH VALVE
MONITOR
CONTROL
REGEN
RELAYS
SERVICE
HELP
Enter
oo
®
ON-BOARD
10 = TIME DELAY
MONITOR
CONTROL
REGEN
RELAYS
SERVICE
HELP
Enter
oo
®
ON-BOARD
-
C
TI
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
3-61