6 Advanced Setting
6.1 Enter the setting
6.1.1 Power on the machine while Holding the
START button and hold for 5seconds. Thiw
will enter into the advanced settings mode.
6.1.2 Select the state (State 1 thru. 3) by pressing
the Program button. Press the START
button to enter the setting.
6.2 S1 state
If you select the S1. You may change
the unit of temperature and pressure,
and adjust time and date.
6.2.1 The first option is to select the unit of
temperature. Press temperature button
℃
.
to select
or
F
The unit you selected will be lighted. Press
the program button to the next item.
6.2.2 You may select the unit of pressure in the same manner.
6.2.3 Then press program button to the next item to adjust the time and
date. After the last word of the date or time is set , then the data is
permitted to be saved. If you want to finish the setting you shall press
START. It will return to the screen of selecting states.
6.3 S2 state
6.3.1 You may check the count of sterilization
cycle. It can not be changed by operator.
6.3.2 Set the parameter for high altitude.
If you have trouble completing a cycle in a
location of high altitude (above 2.0 km or
atmospheric pressure is below 80kPa)
you may need to adjust this parameter.
6.3.3 Language set:
00 English 01 German 02 Spanish
03 Polish
04 French
05 Magyar
06 Romanian 07 Dutch 08 Lithuanian 09 Latvian
The Machine No. And cycle No can not be set by the operator.
11
Altitude set
cycle No
Language
machine No
set
Instructions manual
6.4 S3 state
6.4.1 Adjust the length of sterilization and drying time.
Press program button to select the program.(
Press temperature button to select the temperature of program.
Then press START to adjust the drying time and holding time.
6.4.2 First to adjust the holding time.
Press temperature button to adjust the data.
Press the program button to select
the items.
6.4.3 Press START to save .
6.4.4 Drying time is 0-19.
Holding time of 250 is 1-59.
F
Holding time of 274 is 1-19.
F
Notice: the default sterilization parameters have been chosen to
provide optimal sterilization results. We do not suggest adjusting
these parameters unless it is necessary.
6.5 Printer (Optional)
6.5.1 Connect the printer cable.
6.5.2 Connect the printer power.
Instructions manual
274 F
holding time
drying time
)
1 2
Need help?
Do you have a question about the Bioclave and is the answer not in the manual?