7.6 Replacement of the door seal
7.6.1 Open the chamber door.
7.6.2 Remove the door seal ring carefully by hand.
7.6.3 Clean the door seal ring carefully with a smooth cloth
saturated with distilled water.
7.6.4 Moisten the new seal with medical disinfectant or
isopropyl alcohol.
7.6.5 Insert the new seal and press in sequence as follows:
1). Press in the top and bottom
of the seal ring.
3). Press the remaining sections
of the seal ring.
Caution: Please ensure the chamber and the door have cooled
prior to replacing the seal ring.
1 7
2). Press in the left and right sides
of the seal ring.
4). Press all areas of the seal
surface for a smooth finish.
Instructions manual
7.7 The drain valve
1.Press the included hose on to the
drain valve firmly.
3.Pull the drain valve outward, the
tank will begin to drain.
8 Transportation and Storage
8.1 Switch off the sterilizer before transportation or storage. Pull out the
plug to let the machine cool down.
8.2 Drain the distilled water tank and the used water tank
8.3 Conditions for transportation and storage:
Temperature: -20 ~ +55
Relative humidity: 85%
≥
Atmospheric pressure: 50kPa~106kPa
Instructions manual
2.Set the drain valve to the open
position by turning it counter
clockwise.
4.After finish draining the tank, push
the drain valve inward and turn
clockwise into the closed position .
1 8