7 Maintenance

7. Maintenance
Frequency
Maintenance Operation
Clean the door seal
Daily
Clean the external surface
Clean the distilled water tank
Weekly
Clean the sterilization chamber
Replace the door seal
Every year
7.1 Clean the distilled water tank every
week with isopropyl alcohol or a
medical disinfectant.
7.2 Clean the chamber weekly.
7.2.1 Remove all trays and the tray rack
from the chamber.
7.2.2 Clean the chamber with a
smooth cloth saturated with
distilled water.
7.2.3 Apply the same procedure for
the trays and rack.
1 5
Instructions manual
7.4 Clean the door seal
Clean the door seal weekly, with a smooth cloth saturated with the
distilled water.
7.5 Door adjustment
On normal circumstance the chamber door does not require adjustments.
However, if the seal fails (resulting in steam leaking from the front of
the chamber), you may use the included tool to tighten the door seal.
7.5.1 Open the door
7.5.2 Insert the spanner tool in the gap beneath the plastic cover; use the spanner
to grip the adjusting nut (Fig 1). Turn the nut counter clockwise as
the figure below (Fig 2). This will tighten the sealing plate.
7.5.3 Turn the nut until the sealing plate is tight. If the door knob is too tight,
you may also turn the nut clockwise to loosen it.
Fig 1
Caution: Never adjust the chamber door while the door is closed
Instructions manual
Fig 2
.
1 6