Connecting the Test Station Power Cable

Once all cables and lines have been connected to the modular power supplies and the ARES-G2 test sta-
tion, connect the power cable.
Install the power cable as follows:
1
Make sure the power switch on the lower power supply module is in the Off (O) position.
2
Plug the large, multi-pin cable from the lower power supply to the appropriate connection on the rear
of the ARES-G2 test station, as shown in the figure below.
3
Connect the grounding wire to the cabinet, as shown in the figure below.
4
Plug the power cord from the lower power supply module into the wall outlet.
CAUTION: Before plugging the power cable into the wall outlet, make sure the instrument is com-
patible with the line voltage. Check the label on the back of the unit to verify the voltage.
MISE EN GARDE: Avant de brancher le câble d'alimentation du dans la prise murale, assurez-vous
que l'instrument est compatible avec la tension de la ligne. Consultez l'étiquette au dos de l'appareil
pour vérifier la tension.
Grounding
Power connection
wire
from lower power supply
Figure 38 Power and grounding wire connections on rear of ARES-G2 test station.
ARES-G2 Getting Started Guide
Page 59