Table 4:
Peltier Plate Specifications
Temperature Range
Heating Rate
Precision
Stability
Geometry Compatibility
Optional Accessory
Table 5:
Peltier Plate Cooling Options
ARES Computer-Controlled Circulator
Peltier Circulator Model 612
AR Cooled Circulator Model 340
ARES Peltier Circulator*
*Operating the Peltier Plate at temperatures that are –20 °C or colder depends on maintaining cool water in the circulator reservoir.
Using room temperature water in the reservoir, the Peltier Plate Accessory can hold at –30°C for about 30 minutes, but the Peltier
Plate temperature will drift upwards as the water in the reservoir becomes warmer.
The Peltier Plate may lose temperature control when the coolant reservoir temperature is more than 45°C to 50°C warmer than the
commanded plate temperature.
ARES-G2 Getting Started Guide
–40°C to 180°C, depending on circulator fluid
temperature.
up to 30°C/min
+0.05°C
+0.1°C
All ARES-G2 upper Cone and Plate geometries (up to 50 mm diameter)
Solvent Trap
Option
Temperature Range
–40°C to 180°C
–40°C to 160°C
–20°C to 150°C
–30°C to 140°C
Page 29