C Kv/Ma Settings For Panoramic; Changing Binnings - Planmeca Proline XC Panoramic Calibration Manual

Table of Contents

Advertisement

NOTE:
Make sure the red calibration block is placed onto the collimator before calibrating. (Figure
33)

3.1.c kV/mA settings for Panoramic

3.1.d Changing Binnings
Click Settings and select Binning then the appropriate binning.
(Figure 34)
Press the Ready icon after turning the radiation off or on. Take an
exposure.
NOTE: Make sure when checking all binnings the correct kV/mA are being used.
Image will come out as showing a grey box with a black stripe in the
center. (Figure 35)
NOTE: There is no need to close and re-open Dimax3 Tool to check all binnings. Follow
1.3.a.iiii on pg. 23 for checking the other binnings.
Figure 33
BINNING
kV
mA
4x4
60
3x3
66
2x2
70
Figure 34
Figure 35