Because of the capacity consumption during the storage and transport course, the battery
capacity is not full when it is used for the first time. Battery recharge should be considered
before the first use.
NOTE: The battery will automatically stop charging if you print an ECG report.
Repeated undercharging of the battery will damage the battery and reduce battery life.
3) Replacement
When the useful life of the battery is over, or foul smell and leakage are found, please
contact the manufacturer or the local distributor for replacement.
1. Only qualified service engineers authorized by the manufacturer can open the
battery compartment and replace the battery, and the battery of the same model and
specification provided by the manufacturer must be used.
2. Danger of explosion -- Do not reverse the anode and the cathode when installing the
battery.
3. When the battery's useful life is over, contact the manufacturer or the local distributor
for disposal or dispose of the battery according to local regulations.
4. Remove the battery from the electrocardiograph when the electrocardiograph isn't
used for a long time.
5. If the battery is stored alone and not used for a long time, we recommend that the
battery be charged at least once every 6 months to prevent overdischarge.
If the battery has been fully charged and requires recharging after printing only a few
ECGs, consider replacement.

14.4.2 Recorder Paper

NOTE: Recorder paper provided by the manufacturer should be used. Other paper may
shorten the life of the thermal print head. The deteriorated print head may lead to
illegible ECG reports and block the advance of the paper.
Storage Requirements:
 Recorder paper should be stored in a dry, dark and cool area, avoiding excessive
temperature, humidity and sunshine.
Do not put the recorder paper under fluorescence for a long time.