Chapter 12 Error Messages

Hint information and the corresponding causes provided by the electrocardiograph are listed in
Table 12-1.
Message
Lead off
Battery Weak
No Paper
Testing
Paper Error
Testing
Sampling/Analyzing/Recor
ding
Learning
Detecting
Transmitting
Transmitting fails
Loading Order...
Memory Full
Module Error
DEMO
Overload
Table 12-1 Error Messages and Causes for Resting ECG
Cause
Electrodes fall off the patient or the patient cable falls off the unit,
or a high polarization voltage occurs.
The battery is weak.
Recorder paper runs out or is not loaded.
The ECG data is being sampled periodically.
When Paper Marker is set to Yes, the electrocardiograph advances
the recorder paper to the next black marker. If it advances the paper
for 300mm and cannot find the next black marker, the hint Paper
Error is displayed.
The ECG data is being sampled periodically.
ECG signals are being sampled / analyzed / recorded.
The self-study process of arrhythmia arithmetic in the Trigger
Sample mode
The examining process of arrhythmia data in the Trigger Sample
mode
ECG data is being transmitted from the electrocardiograph to the
PC through the net or serial cable in the auto or rhythm mode.
Data fails to be transmitted through Ethernet or WIFI.
Orders are being loaded to the electrocardiograph.
There is no space for saving more records.
There is something wrong with the signal sample module.
The system is in the demonstration mode.
The direct current offset voltage on an electrode is too high.