High-speed digitizers hardware extension of keysight 89600 vsa software
Hide thumbs Also See for M9703B:

Advertisement

Configure extension parameters for specific acquisition
modes
Interleaved Mode
The option of combining two channels of the digitizer in interleaved mode is available
if your product has been ordered with the -INT option.
Interleaving allows to double the sampling rate with the trade-off of halving the
available center frequency range, as shown in the table below:
Model Option Mode
M9703B -SR1
M9703B Hardware Extension of 89600 VSA Software
Normal (non-inter-
leaved)
-SR2
Normal (non-inter-
leaved)
-SR1
Interleaved
-SR2
Interleaved
For best measurement fidelity, you should ensure that the signal
center frequency + (analog bandwidth/2) does not cross the
Nyquist frequency.
89600 VSA Software Configuration
Center Frequency
Range
0 - 1 GHz
0 - 1.6 GHz
0 - 1.4 GHz
0 - 1.4 GHz
ADC Sampling
Rate
1 GS/s
1.6 GS/s
2 GS/s
3.2 GS/s