5.
Fill the filling syringe with air
and purge the clean water to dry
the vasculature.
Finger Stick Bleed
The left middle finger can bleed real fluid through a precut opening on
the fingertip. The feature allows participants to practice finger stick
technique and collect simulated blood samples.
WARNING: Use only Gaumard's provided simulated
blood to fill the internal reservoirs. Any other simulated blood
brand containing sugar or any additive may cause blockage
and/or interruption of the vasculature system.
Always flush internal fluids with clean water then air to prevent mold.
Avoid overfilling the fluid reservoir. Doing so
may result in a leak at the bleeding site.
Feature
Finger Stick
Instructions for Use
1.
Power the simulator ON
2. Locate the black port on the
simulator's left arm
Using SUSIE® S2000
32 |
Reservoir Capacity
2 cc