Electrical Therapy

For simulations that incorporate electrical therapy of any kind,
always know your equipment, and follow the device manufacturers'
safety guidelines. Always treat the simulator as a real patient.
SUSIE S2000 is equipped with 4 ECG
GREEN
electrode sites marked
generate an ECG lead II waveform
detectable on most of today's ECG
monitors.
Defibrillate only on the large sternum
and apex sites, circled RED.
WARNING: NEVER deliver a shock to ECG electrode targets marked
Green. Doing so will not create a fire hazard, nor is there risk of shock
to the provider, but internal damage to the simulator will result. This
type of damage is considered improper use and it is NOT covered by
the simulator's warranty. The system will require repair at our facility.
SUSIE S2000 can be shocked or paced with real energy for cardioversion,
defibrillation and pacing drills. AED will display SUSIE's ECG, analyze her
cardiac rhythm and advise action.
Instructions for Use
1.
Turn on the simulator. Refer to the Equipment Set-Up section.
2.
Connect the ECG lead wires on
the ECG sites.
3.
Turn on the ECG monitor.
4.
ECG monitor will display the selected ECG rhythm.
UNI provides a library of 20+
cardiac rhythms and options.
Using SUSIE® S2000
20 |
that
SUSIE®
S2000 |
User Guide