PHYWE Cobra4 Operating Instructions

Sensor-unit colorimeter

Advertisement

Quick Links

PHYWE Systeme GmbH & Co. KG
Robert-Bosch-Breite 10
D–37079 Göttingen
Telefon
+49 (0) 551 604–0
Fax
+49 (0) 551 604–107
E-mail
info@phywe.de
Operating instructions
The unit complies
with the correspond-
ing EC guidelines.
TABLE OF CONTENTS
1
SAFETY INSTRUCTIONS
2
PUPOSE AND CHARACTERISTICS
3
FUNCTIONAL AND OPERATING ELEMENTS
4
OPERATING INSTRUCTIONS
5
HANDLING
6
TECHNICAL DATA
7
SCOPE OF SUPPLY
8
ACCESSORIES
9
WARRANTY
10 WASTE DISPOSAL
1
SAFETY INSTRUCTIONS
Read the operating instructions thoroughly and com-
pletely prior to starting the unit. Please read also the op-
erating instructions of the measurement data recording
www.phywe.com, © All rights reserved
Cobra4 Sensor-Unit Colorimeter
Fig. 1: 12634-00 Cobra4 Sensor-Unit Colorimeter
device that is used (Cobra4 Wireless-Link, USB-Link,
Mobile-Link, or Junior-Link). This is important for your
own protection and for avoiding damage to the unit.
Only use the unit for its intended purpose.
The unit is solely intended for use in dry rooms where
there is no risk of explosion.
Protect the Cobra4 Sensor-Unit Colorimeter against dust,
moisture, and vapours. Clean the unit with a slightly
moist, lint-free cloth. Do not use any aggressive cleaning
agents or solvents.
Ensure that no liquid can penetrate the housing. Other-
wise, the Cobra4 Sensor-Unit Colorimeter may be dam-
aged.
When using the unit outdoors, ensure that no rain or
splash water can penetrate the housing.
Do not open the unit.
Connect the data output of the Cobra4 Sensor-Unit Col-
orimeter only to the intended measurement data re-
cording devices.
Based on today's state of software development, it is im-
possible to guarantee that a product is completely free
from errors or defects.
This is why PHYWE Systeme GmbH & Co. KG does not
accept any liability for damage that results from the in-
stallation or use of this unit.
2
PUPOSE AND CHARACTERISTICS
The Cobra4 Sensor-Unit Colorimeter is used for measuring
the absorption or transmission of light. For this purpose, there
are four wavelengths: red: 650 nm, orange: 610 nm, green:
520 nm, and blue: 470 nm.
12634-00
1
12634-00 / 0915

Advertisement

loading
Need help?

Need help?

Do you have a question about the Cobra4 and is the answer not in the manual?

Questions and answers

Summary of Contents for PHYWE Cobra4

  • Page 1 10 WASTE DISPOSAL This is why PHYWE Systeme GmbH & Co. KG does not accept any liability for damage that results from the in- stallation or use of this unit.
  • Page 2 Software/device software update Ensure that the current versions of the "measure" software • Cobra4 Junior-Link (12615-00) for the data transfer via package and of the firmware of the Cobra4 module are in- USB during student experiments, stalled. • hand-held Cobra4 Mobile-Link (12620-10).
  • Page 3 Absorbance (Abs): The absorbance A is calculated based on proper waste disposal. the intensity without the sample (I ) and based on the inten- sity with the sample (I): PHYWE Systeme GmbH & Co. KG Customer Service Robert-Bosch-Breite 10 D–37079 Göttingen It is a unit-free measure.