Keysight N5245-60121 Installation Note page 36

Add 4-port capability upgrade kit

Advertisement

Installation Procedure for the Upgrade
2. Copy the license file(s) from the email(s) to the root d irectory of the USB
3. Insert the USB flash drive into the PNA's USB drive slot.
4. The instrument automatically reads the license file(s) from the USB flash
5. Wait >1 minute for your Option to be installed.
6. The PNA application automatically restarts after the option installation has
Attempting to re-install a license file that is already installed may generate
a "Corrupt Media" error message. Ignore this message.
Option Verification Procedure
Once the analyzer has restarted and the Network Analyzer program is again
running:
1. On the analyzer's Help menu, press About Network Analyzer.
2. Verify that "422" is listed, but "222" is not listed after "Options:" in the
3. Click OK.
If the option has not been enabled or if the older option has not been
removed, contact Keysight Technologies. Refer to
from Keysight" on page
36
flash drive.
More than one license file may be copied to the USB flash drive.
drive, validates the file(s), and installs the option(s). Generally, warnings
or error messages may be ignored.
The instrument will not prompt you when the upgrade has been
completed.
been completed.
Optional: Cycle power on the instrument to activate the newly installed
option upgrade(s). (Press the power button once to power down the
instrument.)
display.
4.
"Getting Assistance
Installation Note N5245-90121