Electrical connection
4.1
Conductor material
•
Power supply cable: flexible standard conductor material
•
Maximum wire cross-section: 1.5 mm
Notice - Potential damage to parts
Using rigid conductor material may cause line breaks.
4.2
Electrical connections configuration
Fig. 6
4.3
Sensor connection
Depending on the sensor model, a variety of conductor materials can be used for sensor
connections. The integrated internal reference junction makes it possible to directly connect
thermal compensating cables.
For the solder terminal, a soldering tag is provided at each terminal.
2
3
1
1
1
2
2
3
3
4
Fig. 7
1
Potentiometer, Four-wire circuit
2
Potentiometer, Three-wire circuit
3
Potentiometer, Two-wire circuit
14 - EN
Sensor connection
4
5
1
1
1
2
2
J
J
J
3
3
3
4
4
5
6
2
(16 AWG)
1
2
3
4
Interface for HMI LCD display type AS
7
8
6
1
1
1
2
2
3
RTD, Four-wire circuit
RTD, Three-wire circuit
RTD, Two-wire circuit
TTH200
Signal/supply connection
-
1
2
3
4
7
Voltage measurement
8
Thermocouple
CI/TTH200-EN
A00231
-