3.6

Luminance adjustment

To adjust the luminance
1. Bring up the OSD main menu.
2. Navigate to the Adjustments > Luminance menu.
3. Set a luminance value as desired and confirm.
3.7

sRGB color space

About sRGB color space
The sRGB color space combines a display function and white point selection and is designed to match
typical home and office viewing conditions. It is widely used in most computer applications.
When selecting sRGB, the Display function and White point selection options in the Ad-
justments menu will be disabled.
To select sRGB color space
1. Bring up the OSD main menu.
2. Navigate to the Adjustments menu.
3. Select sRGB and confirm.
3.8

Display functions

Display function selection is disabled when sRGB is selected in the Adjustments menu.
About display functions
Native, uncorrected panels will display all grayscale/color levels with luminance increments that are not
optimal for crucial diagnostic information. Studies have shown however, that in medical images certain
grayscale/color parts contain more diagnostic information then others. To respond to these conclusions,
display functions have been defined. These functions emphasize on these parts containing crucial diag-
nostic information by correcting the native panel behavior.
The available display functions for your Nio Color 2MP 23" are:
Native: If you select Native, the native panel behavior will not be corrected.
DICOM: DICOM (Digital Imaging and Communications in Medicine) is an international standard that
was developed to improve the quality and communication of digital images in radiology. In short, the
DICOM display function results in more visible grayscales in the images. Barco recommends selecting
the DICOM display function for most medical viewing applications.
The DICOM display function applies ambient light compensation (ALC) taking the ambient light con-
ditions of your reading room into account. The available reading room options are:
- Darkroom: Selects DICOM calibrated function, optimized for darkroom conditions (0 Lux)
- Office: Selects DICOM function optimized for office conditions (60-180 Lux)
- Operation Room: Selects DICOM function optimized for operating room conditions (300-400 Lux)
K5902134 NIO COLOR 2MP 23" 05/07/2017